The PREVAC EA15-HP1 hemispherical energy analyser, which is based on the EA15 analyser, allows due to special designed extraction cones and automatic vacuum control, high-resolution PES measurements in static and dynamic environments up to 1 mbar.
Equipped with a total number of 11 slits, the analyser offers the possibility to choose between best energy resolution and best intensity. According to given photoelectron energy the analyser is set up with up to 8 predefined PE to satisfy customer’s requirements.
– 40 mm diameter dual MCP detector
– 656 energy channels available simultanously
– 494 angular channels available simultanously
– 90 fps
The analyser package includes:
EA15-HP1 hemispherical energy analyser can be controlled directly by LabVIEW programming environment.
Mounting flange DN 100 CF
Bakeout temperature up to 120 °C
Working distance 1-2 mm (dependent on cone opening)
Analyser mean radius 150 mm
Pass energies 1, 2, 5, 10, 20, 50, 100, 200 eV
Energy resolution
XPS < 20 meV FWHM
XPS/UPS & XPS/UPS/ARPES < 3 meV FWHM
Kinetic energy range 0.5 – 2000eV
Acquisition modes fixed, scan
Transmission and angular lens mode:
Maximum energy 12.5 % of pass energy
window in fixed mode (for CCD-MCD)
Completely designed of non-magnetic materials
NOTE | The use of a mu-metal analysis vacuum chamber or an inner mu-metal shielded chamber in combination with PREVAC EA15-HP1 analyser is recommended in order to minimalize influence of external magnetic fields and to maximize performance of the analyser.